The silver activation counters are commonly used for pulsed-neutron yield measurements especially in plasma focus devices. The counter normally consists of a Geiger-Muller tube along with silver foils and polyethylene (as a moderator), which is calibrated against an Am-Be radioisotope neutron source. The neutrons, after being slowed-down in the polyethylene, activate the silver foils. By measuring the foil activity with a Geiger-Muller counter, the neutron yield is determined. In the present paper, the activation counter’s calibration constant calculation using the MCNP4C code is explained. The calculated calibration constant is in good agreement with the experimental results.
Etaati,G. , Amrollahi,R. and Doostmohammadi,V. (2019). Calibration constant of a silver activation counter used with plasma focus devices. Iranian Journal of Physics Research, 9(4), 305-312.
MLA
Etaati,G. , , Amrollahi,R. , and Doostmohammadi,V. . "Calibration constant of a silver activation counter used with plasma focus devices", Iranian Journal of Physics Research, 9, 4, 2019, 305-312.
HARVARD
Etaati G., Amrollahi R., Doostmohammadi V. (2019). 'Calibration constant of a silver activation counter used with plasma focus devices', Iranian Journal of Physics Research, 9(4), pp. 305-312.
CHICAGO
G. Etaati, R. Amrollahi and V. Doostmohammadi, "Calibration constant of a silver activation counter used with plasma focus devices," Iranian Journal of Physics Research, 9 4 (2019): 305-312,
VANCOUVER
Etaati G., Amrollahi R., Doostmohammadi V. Calibration constant of a silver activation counter used with plasma focus devices. Dear user; Recently we have changed our software to Sinaweb. If you had already registered with the old site, you may use the same USERNAME but you need to change your password. To do so at the first use, please choose, 2019; 9(4): 305-312.