In this paper a one dimentional position-sensitive X-ray gas detector has been designed, simulated and constructed based on Multi Wire Proportional Chamber (MWPC) with delay line position readout. These kinds of detectors are useful in soft X-ray imaging and are capable of being extended to two dimensions position readout easily. The position resolution of this detector is estimated to be 230µm
Karami,H , Jafarzadeh Khatibani,M , Rahighi,J and Kargar,Z . (2019). Design, simulation and construction of a position sensitive X-ray gas detector. Iranian Journal of Physics Research, 17(2), 299-306.
MLA
Karami,H , , Jafarzadeh Khatibani,M , , Rahighi,J , and Kargar,Z . "Design, simulation and construction of a position sensitive X-ray gas detector", Iranian Journal of Physics Research, 17, 2, 2019, 299-306.
HARVARD
Karami H, Jafarzadeh Khatibani M, Rahighi J, Kargar Z. (2019). 'Design, simulation and construction of a position sensitive X-ray gas detector', Iranian Journal of Physics Research, 17(2), pp. 299-306.
CHICAGO
H Karami, M Jafarzadeh Khatibani, J Rahighi and Z Kargar, "Design, simulation and construction of a position sensitive X-ray gas detector," Iranian Journal of Physics Research, 17 2 (2019): 299-306,
VANCOUVER
Karami H, Jafarzadeh Khatibani M, Rahighi J, Kargar Z. Design, simulation and construction of a position sensitive X-ray gas detector. Iranian Journal of Physics Research. 2019;17(2):299-306 (In Persian).