Abstract

Interest in slot scanning imaging system has been increased recently because of its advantages such as high image resolution, reduction in patient dose and reduction in image distortion. The purpose of this study is to simulate a slot scanning imaging system using GATE multi-purpose Monte Carlo code and design an object phantom to evaluate image quality. The energy spectrum of X-ray in the radiological diagnostic range was simulated by GATE Monte Carlo code. An object phantom consisting of stripes with different thicknesses of copper was then designed to determine CNR and line pairs per millimeter to determine spatial resolution.
The results showed that the simulated energy spectrum were in accordance with experiment. The evaluated parameters of image quality obtained from the simulations were compared with X-ray images using standard image quality criteria to determine the capabilities of the slot scanning imaging method. The spatial resolution of the simulated images was obtained to be 1.6 lp / mm for the slot scanning method, which is in the range of clinical radiology images. Therefore, this simulated model can be used to investigate and optimize other influential parameters such as filtration effect and collimator slit width effect, object distance to the detector ... in slot scanning imaging system.

Keywords

  1. T R Fewell and R E Shuping, Phys. 4 (1977) 187.

  2. D R Dance, A Thilander Klang, M Sandborg, C L Skinner, I A Castellano Smith, and G Alm Carlsson, Br J Radiol 73 (2000) 1056.

  3. L J M Kroft, V J Geleijns, B J A Mertens, W J H Veldkamp, H M Zonderland, and A de Roos, Radiology 231 (????)

  4. M Zähringer, B Krug, K F Kamm, G Wassmer, M Hellmich, G Winnekendonk, J Andermahr, A Goßmann, and K J Lackner. AJR Am J Roentgenol. 177 (2001) 1397.

  5. D W Piraino, W J Davros, M Lieber, B J Richmond, J R Schils, M P Recht, P N Grooff, G H Belhobek. AJR Am J Roentgenol. 172 (1999) 177.

  6. K H Lee, J W Kwon, Y Ch Yoon, S Choi, J Y Jung, J H Kim, S J Lee, Korean Journal of Radiology 10 (1) (2009) 51

  7. T D Perks, R Dendere, B J Irving, T Hartley, P Scholtz, A Lawson, C Trauernicht, S Steiner, and T S Douglas. Radiation Protection Dosimetry, 1 (2014).

  8. M Kulkarni, R Dendere, F Nicolls, S Steiner, T S Douglas, Physica Medica 32 (2016) 284.

  9. J M Boone, K K Lindfors, N II Cooper, and J A Seibert. Med Phys 27 (2000) 2408.

  10. K Viswanathan, K Balasubramaniam. Proceedings of the National Seminar & Exhibition on Non-Destructive Evaluation, NDE. India: December (2009) 10-12.

  11. U Pietrzyk, A Zakhnini, M Axer, S Sauerzapf, D Benoit, M Gaens. Phys. 23 (2013) 65.

  12. B J Irving, G J Maree, E R Hering, and T S Douglas. Radiation Protection Dosimetry, 130 (4) (2008) 482.

  13. M T Bahreyni‑Toosi, SH Nasseri, M Momennezhad, F Hasanabadi, H Gholamhosseinian. Journal of Medical Signals & Sensors, 4 (2014) 268.

  14. M Scheelke, JH Potgieter, M de Villiers. In: Proc. SPIE, SPIE Medical Imaging: Physics of Medical Imaging. 5745 (2005).

  15. N Banihashemi, J Soltani-Nabipour, A Khorshidi, and H Mohammadi. Phys. J. Plus (2020) 135:269

  16. M De Villiers, and M De Jager. In: Proc. SPIE, 5030. Medical Imaging. (2003).

  17. International Commission on Radiation Units and Measurements (ICRU). ICRU; (1996) ICRU Report 54.

  18. L C d S Romualdo, M A d C Vieira and H Schiabel. XXII Brazilian Symposium on Computer Graphics and Image Processing, Rio de Janiero, (2009) 180.

  19. E Mah, E Samei, D Peck, Journal of Applied Clinical Medical Physics, 2, 2, Spring (2001).

  20. M R Ay, M Shahriari, S Sarkar, M Adib, and H Zaidi. Phys Med Biol 49 (2004) 4897.

  21. H Mohammadi, M R Shojaei, J Soltani-Nabipour. Applied Radiation and Isotopes 171 (2021) 109642.


 

تحت نظارت وف ایرانی